PARTICLE MASS COUNTER / EN
09/15/17 Version No. 00
19
To set the K-Factor coefficients:
1. Press
F2
(SETUP) key on the Main screen. The Setup Page 1 screen is
displayed.
2. Press
F2
(PAGE 2) key on the Setup Page 1 screen. The Setup Page 2
screen is displayed.
3. Press
F2
(PAGE 3) key on the Setup Page 2 screen. The Setup Page 3
screen is displayed.
4. Press
or
key to move the cursor to the desired measurement
range.
5. Press
F1
(EDIT) key to enter the edit mode.
6. Press
or
key to setting the desired coefficient value from 0.1 to
9.99 (default value is 1.00).
7. Once the desired value has been selected, press
F1
(SAVE) key.
8. Press
(RETURN) key to return to the Main screen.
7-2 Counter Operation
Once the meter configuration is set:
Use the zero count filter to purging the meter.
Collect the air samples.
Transfer the data from the air sample to PC for further analysis.
Review the number of samples in the buffer.
Clear the buffer if necessary.