Heavy Duty True RMS Digital Multimeter/English

Heavy Duty True RMS Digital Multimeter/English

25

13/08/2017 Version No. 001

13/08/2017 Version No. 001

7-12 Li-ion Battery Charge

2.Insert the socket into the Meter Input port. And the Adapter connected to the switch 

socket.
Then Insert the Adapter into Power socket.

Enclosure
Shock (Drop Test)
Diode Test

Continuity Check

PEAK
Temperature Sensor
Input Impedance
AC Response

Double molded, waterproof

Test cur

Audible signal will sound if the resistance is less 
than 25

Ω

 (approx.), test cur

R

mocouple

AC

True rms

Adapter 

True RMS

Multimeter

 

 

F1

HOLD

REL

RANGE

F2

F3

F4

Bluetooth

5

20A for 

30 sec

MAX 

every

15 min

TEM P

C A P

AC True RMS

ACV Bandwidth
Display
Overrange indication
Auto Power Off
Polarity

Measurement Rate
Low Battery Indication

Battery
Fuses

Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Operating Altitude
Safety

The term stands for “Root-Mean-Square,” which 
represents the method of calculation of the voltage 
or current value. Average responding multimeters 
are calibrated to read correctly only on sine waves 
and they will read inaccurately on non-sine wave or 
distorted signals. True rms meters read accurately 
on either type of signal.
50Hz to 20000Hz
50,000 count TFTLCD
“OL” is displayed
5-30minutes (approximately) with disable feature
Automatic (no indication for positive); Minus (-) 
sign for negative 
20 times per second, nominal 
“    ” is displayed if battery voltage drops below 
operating voltage
One 7.2 volt (NEDA 1604) battery
mA, µA ranges; 0.8A/1000V ceramic fast blow A 
range; 10A/1000V ceramic fast blow
5ºC to 40ºC (41ºF to 104ºF)
-20ºC to 60ºC

 (-4ºF to 140ºF)

Max 80% up to 87ºF

 

(31ºC) decreasing linearly to 

50% at 40ºC (104ºF)
<80% 7000ft. (2000meters) maximum.
This meter is intended for origin of installation use 
and protected, against the 
users, by double insulation per EN61010-1 
and IEC61010-1 2nd Edition (2001) to 
Category IV 600V and Category III 1000V; Pollution 
Degree 2. The meter also meets UL 61010-1, 2nd 
Edition (2004), CAN/CSA C22.2 No. 61010-1 2nd 
Edition (2004),and UL 61010B-2-031, 1st Edition (2003)

Summary of Contents for RS-989

Page 1: ...Centre Phase 2 No 618 Yan an Eastern Road Shanghai 200001 China www rs components com Europe RS Components Ltd PO Box 99 Corby Northants NN17 9RS United Kingdom www rs components com Japan RS Compone...

Page 2: ...e Proper use and care of this meter will provide many years of reliable service 2 Safety WARNING CAUTION MAX This symbol adjacent to another symbol terminal or operating device indicates that the oper...

Page 3: ...quipment Input Protection Limits Function V DC or V AC mA AC DC A AC DC Frequency Resistance Capacitance Duty Cycle Diode Test Continuity Temperature Surge Protection 8kV peak per IEC 61010 Maximum In...

Page 4: ...button for greater that 1 second will return Auto Range HOLD REL Freezes the present reading in the display and allows the display to be saved If Pressing the HOLD REL button for greater that 1 secon...

Page 5: ...nd CONTINUITY measurements Frequency measurements Temperature measurements AC dc amps measurements AC dc milliamps measurements 4 20MA measurements AC dc microampere measurements up to 5 000 uA CAP 4...

Page 6: ...to indicate AC DC on the display 4 Read AC DC measurement value in the display Bluetooth 5 20A for 30 sec MAX every 15 min TEM P C A P 4 2 0 TEM P C F RANGE SETUP SAVE HOLD Auto range 00 00 True RMS M...

Page 7: ...F4 Continuity Check WARNING To avoid electric shock disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements Remove the batteries and unplug the l...

Page 8: ...4 Read the temperature in the display Bluetooth 5 20A for 30 sec MAX every 15 min TEM P C A P 4 2 0 TEM P C F RANGE SETUP SAVE HOLD Auto range 00 00 True RMS Multimeter 0078 0 F F1 HOLD REL RANGE F2 F...

Page 9: ...test lead banana plug into the 10A jack 5 Press the MODE button to indicate AC on the display 6 Read the current in the display 5 13 4 20mA MEASUREMENTS 1 Set up and connect as described for DC mA mea...

Page 10: ...collected data Pressing the softkey labeled Close exits the MIN MAX record session without saving the collected data To save the MIN MAX screen data the MIN MAX session must be ended by pressing the...

Page 11: ...F1 V 0 3078 06 02 14 06 32pm PREV 0 1 2 3 4 View Men 06 02 14 06 32pm Rate 0 2 s Range 50M ViewGrap ENTER PREV CL 5V Press the and soft keys to move cursor Press the softkey labeled Then pressing the...

Page 12: ...CLOSE Fno Graph According to graphically display measurement information Fno Mode Releted to the rotary switch function Fno Save Accesses the memory management menu for saving measurement setting up...

Page 13: ...Bluetooth You can use the Bluetooth communication link and transfer the contents of a meter s memory to a PC Open the setup menu Position the menu selector next to the menu item labeled Bluetooth and...

Page 14: ...uare which represents the method of calculation of the voltage or current value Average responding multimeters are calibrated to read correctly only on sine waves and they will read inaccurately on no...

Page 15: ...g time reading will increase 2 C 500 A 5000 A 50mA 500mA 10A 500 A 5000 A 50mA 500mA 10A 20A 30 sec max with reduced accuracy All AC current ranges are specified from 5 of range to 100 of range 5nF 1...

Page 16: ...in 20 to 80 duty cycle and 100kHz 5Vrms min 20 to 80 duty cycle and 100kHz 10 00Hz 10KHz Sensitivity 2Vrms 0 1 to 99 90 Pulse width 100 s 100ms Frequency 5Hz to 150kHz Resolution 0 001Hz 0 01Hz 0 0001...