12
Heavy Duty True RMS Digital Multimeter/English
Heavy Duty True RMS Digital Multimeter/English
13
13/08/2017 Version No. 001
13/08/2017 Version No. 001
emperature Measurements
Bluetooth
5
20A for
30 sec
MAX
every
15 min
TEMP
C A P
4 ~ 2 0
%
TE M P
°C °F
RANGE SETUP SAVE HOLD
Auto range
00:00
True RMS Multimeter
0
20. 0
0
µF
WARNING:
To avoid electric shock, disconnect power to the unit under test
and discharge all capacitors before taking any capacitance
measurements. Remove the batteries and unplug the line cords.
1.Set the rotary function switch to the green
Ω CAP
position.
2. Press the softkey labeled Mode. Switch to
CAP
3.Insert the black test lead banana plug into the
negative
COM
jack. Insert the red test lead banana
plug into the positive
V
jack.
4.Read the capacitance value in the Display
-
. T
1.Set the function switch to the green
TEMP(°C
or °F)
position.
2.Press the softkey labeled Mode. Switch
TEMP
(
°C or °F
).
3.Insert the Temperature Probe into the input jacks,
making sure to observe the correct polarity.
4.Read the temperature in the display
Bluetooth
5
20A for
30 sec
MAX
every
15 min
TE M P
C A P
4 ~ 2 0
%
TE M P
°C °F
RANGE SETUP SAVE HOLD
Auto range
00:00
True RMS Multimeter
00
78 0
.
°F
F1
HOLD
REL
RANGE
F2
F3
F4
F1
HOLD
REL
RANGE
F2
F3
F4
Bluetooth
5
20A for
30 sec
MAX
every
15 min
TE M P
C A P
4 ~ 2 0
%
TE M P
°C °F
RANGE SETUP SAVE HOLD
Auto range
00:00
True RMS Multimeter
1 0
0. 00
A~
0
1
2
3
4
5
5-11. DC Current Measurements
CAUTION:
Do not make 20A current measurements for longer than 30
seconds.
Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1.Insert the black test lead banana plug into the
negative
COM
jack.
2.For current measurements up to 5000µA DC,
set the function switch to the yellow
µA
position
and insert the red test lead banana plug into the
µA/mA
jack.
3.For current measurements up to500mA DC,
set the function switch to the yellow mA position
and insert the red test lead banana plug into the
µA/mA jack.
4.For current measurements up to 10A DC, set
the function switch to the yellow 10A position
and insert the red test lead banana plug into the
10A jack.
5.Press the
MODE
button to indicate “DC” on the
display.
6.Read the current in the display.
F1
HOLD
REL
RANGE
F2
F3
F4
A~