14
15
Range
Resolution
Accuracy
600 mV
0.1 mV
±0.1% of rdg ±2 digits
6 V
1 mV
60 V
10 mV
600 V
100 mV
1000 V
1V
±0.3% of rdg ±2 digits
Range
Resolution
Accuracy
6 V
1 mV
±0.8% of rdg ±4 digits
60 V
10 mV
600 V
100 mV
1000 V
1 V
±1.2% of rdg ±4 digits
Range
Resolution
Accuracy
600 uA
0.1 uA
±0.8% of rdg ±3 digits
6000 uA
1 uA
60 mA
10 uA
600 mA
100 uA
±
1.2% of rdg
±
3 digits
10 A
10m mA
±
1.8% of rdg
±
3 digits
Range
Resolution
Accuracy
600.0 uA
0.1 uA
±1.0% of rdg ±3 digits
6000 uA
1 uA
60.00 mA
10 uA
600.0 mA
100 uA
±
1.2% of rdg
±
3 digits
10 A
10 mA
±
2.0% of rdg
±
3 digits
Range
Resolution
Accuracy
600.0 Ω
0.1 Ω
±0.5% of rdg ±4 digits
6.000 KΩ
1 Ω
60.00 KΩ
10 Ω
±0.5% of rdg ±2 digits
600.0 KΩ
100 Ω
6.00 MΩ
1 KΩ
±1.5% of rdg ±8 digits
60.00 MΩ
10 KΩ
Range
Resolution
Accuracy
40 nF
10 pF
±5.0% of rdg ±20 digits
400 nF
0.1nF
4 uF
1nF
±3.0% of rdg ±5 digits
40 uF
10 nF
400 uF
0.1uF
±5.0% of rdg ±10 digits
4000 uF
1uF
13/12/2017 Version No. 001
13/12/2017 Version No. 001
11.Specifications
DC Voltage (Auto-ranging)
AC Voltage (Auto-ranging)
DC Current (Auto-ranging)
AC Current (Auto-ranging)
Resistance [ Ω ] (Auto-ranging)
Capacitance (Auto-ranging)
Input Impedance: 10M .
Maximum Input: 1000V dc or 1000V ac rms.
All AC voltage ranges are specified from 5% of range to 100% of range
Input Impedance: 10M .
AC Response: 50Hz to 400Hz
Maximum Input: 1000V dc or 1000V ac rms.
Overload Protection: FF800mA 1000V and F10A 1000V Fuse.
Maximum Input: 6000uA dc on uA range
800mA dc on mA range
10A dc on 10A range.
All AC Current ranges are specified from 5% of range to 100% of range
Overload Protection:FF800mA/1000V and F10A/1000V Fuse.
AC Response: 50 Hz to 400 Hz
Maximum Input: 6000uA ac rms on uA
800mA ac rms on mA
10A ac rms on 10A range.
Input Protection: 1000V dc or 1000V ac rms.
Input Protection: 1000V dc or 1000V ac rms.
Digital Multimeter/ English
Digital Multimeter/ English