IDM505
Range
Frequency
Accuracy
100.000mV
DC
–
± (0.025% + 40d)
1000.00mV
DC
± (0.020% + 20d)
10.0000V
DC
100.000V
DC
1000.00V
DC
± (0.015% + 20d)
100.000mV
AC
1000.00mV
AC
40Hz ~ 65Hz
± (0.70% + 50d)
[1]
66Hz ~ 1KHz
± (1.50% + 50d)
[1]
1.01KHz ~ 3KHz
± (3.00% + 50d)
[2]
10.0000V
AC
100.000V
AC
1000.00V
AC
40Hz ~ 45Hz
± (1.00% + 50d)
[1]
46Hz ~ 65Hz
± (0.40% + 50d)
[1]
66Hz ~ 1KHz
± (1.00% + 50d)
[1]
10.0000V
AC
100.000V
AC
1.01KHz ~ 10KHz
± (2.00% + 50d)
[1]
10.01KHz ~ 20KHz
± (3.00% + 50d)
[2]
20.01KHz ~ 50KHz
± (5.00% + 50d)
[3]
50.01KHz ~ 100KHz
± (10.0% + 50d)
[3] [4]
[1] Below 5% of range, add 70 counts. Below 45Hz, < 50dgt rolling.
[2] Below 5% of range, add 150 counts.[4] At 1000.0VAC,
the accuracy is ± (10.0% + 50d).
[3] Below 5% of range, add 350 counts.
[4] At 100.000VAC, the accuracy is ± (15.0% + 50d).