12
Doc. No.
:
P30001PE0101-07
3.5.
Storage Conditions
➢
Temperature
:
+5 ~ +40 deg. C
➢
Relative humidity
:
15 ~ 75 %
※
Non-condensing
➢
Barometric pressure
:
More than 70kPa (min)
3.6.
Transport Conditions
➢
Temperature
:
- 25 ~ + 60 deg. C
➢
Relative humidity
:
15 ~ 75%
※
Non-condensing
➢
Barometric pressure
:
More than 70kPa (min)
3.7.
Environmental Requirements and Instrument life
➢
Overvoltage category
:Ⅱ
Transient overvoltage
:
2500 V
➢
Pollution degree
:
2
➢
Installation site
:
Indoor use only
➢
Temperature
:
+15 ~ +30 deg. C
➢
Relative humidity
:
15 - 75%
※
Non-condensing
➢
Altitude
:
Up to 2,000 m above the sea level
➢
Space
:
Maintain sufficient distances from the walls as indicated below for
ventilation, maintenance work, etc. Also maintain more than 200 mm free space
above the instrument.
➢
Designed to last 5 years or more (4 runs/day, 250 day/year) with proper
maintenance.
Environmental Requirements
magLEAD 12gC
More than 200 mm
Front
Back
Wall
Wall
Wall
More
than 200
mm
More
than 200
mm