Operating instructions for PODOLOG NOVA 3s | REV. 0
5.8
Exchange of alternate chuck
If the function of the alternate chuck is no longer satisfactory, simply exchange the chuck
with the interchangeable chuck for PODOLOG NOVA 3s (art. No. 11067)
To prevent damage to the handpiece motor, follow the exchange procedure
of the chuck of the enclosed instruction manual
6.
MAINTENANCE, CLEANING AND DISINFECTION
6.1 MAINTENANCE BY RUCK SERVICE
6.2
CLEANING AND DISINFECTION
After every treatment
the outer surfaces of the handpiece and front panel of the controller are
wiped with a suitable surface disinfectant.
When work is completed
the handpiece sleeve is removed (5.7.1) and all accessible surfaces
(inside of handpiece, inside and outside of sleeve) are cleaned and wiped with disinfectant.
The suction hose is also wiped with disin-fectant and visually inspected for cracks or damage.
Also check the housing and in particular the glass front for damage.
To disinfect all parts of the controller and handpiece, use RUCK alcohol-free disinfection wipes
(art. no. 2942801). These can be found in our main cata-logue or at www.hellmut-ruck.de.
To ensure continued safe and correct operation of the medical prod-uct, the PO-
DOLOG NOVA 3s must be maintained by RUCK SERVICE every 900 operating
hours or after 24 months at the latest. During maintenance an electrical safety
inspection is carried out in accord-ance with DIN EN 62353.
IF THIS REQUIREMENT IS NOT SATISFIED, THE GUARANTEE WILL LAPSE
AND WE CANNOT ACCEPT LIABILITY!
Risk of disease transmission. Insufficient disinfection can result in transmission of
disease. Always comply with the requirements of the re-sponsible national body
for hygiene and disinfection. Use the RUCK® disinfection cloths – alcohol-free
(art. no. 2942801) – do not use any spray disinfectants. It must be possible to ef-
fectively evaporate surface disinfectants before the next treatment; allow surface
disinfectants to thoroughly dry as specified by the manufacturer, but do not wipe
surfaces dry.