46
Operating instructions for PODOLOG NOVA 3s | REV. 0
Guidelines and manufacturer’s declaration – electromagnetic interference immunity
The PODOLOG NOVA 3s is designed for use in electromagnetic environments as specified
below. The customer or user of the PODOLOG NOVA 3s should ensure that the device is
used in the prescribed environment.
Interference
immunity tests
IEC 60601-
Test
level
Compliance
levell
Guidelines for the
electro-magnetic environment
Conducted HF
in-terference
variables
according to
IEC 61000-4-6
Beamed
HF interfe-
rence variables
accord-ing to
IEC 61000-4-3
3 Veff
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
3 Veff
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
Portable and mobile radio devices
shall not be used within a closer
proximity to the PODOLOG NOVA 3s
Footcare System, including the ca-
bles, than the recommended safety
distance, calculated using the equa-
tion applicable for the trans-mission
frequency.
Recommended safety distance:
d=(3.5/V1)√P
d=(3.5/E1)√P
d=(7/V1)√P
80 MHz to 800 MHz
800 MHz to 2.5 GHz
With P as the nominal power of the
transmitter in watt according to the
information given by the transmitter
manufacturer, and with d as the re-
commended safety distance in
meters (m).