10
appropriately. Repeat the test for many times to obtain the average value of the liquid volume
obtained from multiple calibrations, which can improve the accuracy of the calibration results.
Figure 3-2-2
Editing and process of calibration
3.2.2 Inquiry Mode
Parameters such as standard speed, standard flow, minimum flow, maximum flow and standard
ratio can be queried.
3.2.2.1 Calibrated speed
Last calibrated speed value. The default is 100.0 rpm
3.2.2.2 Standard flow
After the calibration, the system automatically calculates the calibrated flow under the correct
input of the calibrated speed and the liquid volume value under the fixed time. Flow unit: ml / min.
3.2.2.3 Minimum flow
According to the calibrated speed, minimum speed (0.1 rpm) and calibrated flow, the system