CARE, FUSE REPLACEMENT AND SERVICE
Carl Zeiss
Instrument maintenance
Axio Lab.A1
108 430037-7144-001
04/2013
5.2
Instrument maintenance
5.2.1
Checking the instrument
x
Ensure that the prescribed line voltages are observed.
x
Check the power cable and the plug for possible damage.
x
If any damage is observed, turn the instrument off and secure immediately. Call in a qualified
professional to remedy the problem.
5.2.2
Replacing the fuses in the stand
Always disconnect the instrument from
the power supply before replacing
fuses.
If the fuses fail, the reason must first of all be
ascertained and technical problems properly
remedied.
The fuse box is located on the rear side of the
microscope. It is combined with the supply plug
and contains two type
T 3.15 A/H /250 V
fuses.
x
Disconnect from the power supply
x
Remove the fuse holder (Fig. 5-1/
2
) by pulling it
to the front. Use a small screwdriver for the
purpose if necessary.
x
Remove the fuses from the fuse holder and
replace with new fuses.
x
Push the fuse holder back into the fuse box
(Fig. 5-1/
1
) until it engages.
x
Connect the power plug.
Fig. 5-1
Replacing the fuses in the stand