DESCRIPTION OF THE INSTRUMENT
Axio Lab.A1
Technical data
Carl Zeiss
04/2013 430037-7144-001
21
2.3
Technical data
Dimensions (width x depth x height)
Axio Lab.A1 basic microscope stand
without tube (430037-9000-000) ................................................ approx. 219 mm x 410 mm x 3395 mm
The other stand types differ slightly in depth and significantly in height, depending on the tube used. An
overview of viewing heights (eyepoint heights) of the various tubes can be found in Section 2.3.1.
An estimate of the height of the stand with the respective tube can be obtained by adding:
10 mm to the viewing height in the lower position of binocular section
for tubes with a fixed viewing angle
10 mm to the viewing height of the upper limit for ergo tubes
Weight
Axio Lab.A1 microscope stand (depending on version and accessories) ......................... approx. 8 to 20 kg
Ambient conditions
Shipping (in packaging):
Permissible ambient temperature ................................................................................... -40 to +70 °C
Storage:
Permissible ambient temperature .................................................................................. +10 to +40 °C
Permissible humidity (without condensation) ......................................................... max. 75 % at 35 °C
Operation:
Permissible ambient temperature .................................................................................. +10 to +40 °C
Permissible relative humidity (without condensation) ............................................. max. 75 % at 35 °C
Highest permitted altitude of use .................................................................................... max. 2000 m
Air pressure ........................................................................................................800 hPa to 1060 hPa
Degree of pollution .......................................................................................................................... 2
Operational specifications
Operational area ................................................................................................................. Closed rooms
Protective class .......................................................................................................................................I
Protection type ................................................................................................................................ IP 20
Electrical safety .............................................................................. acc. to DIN EN 61010-1 (IEC 61010-1)
in conformance with CSA and UL regulations
Overvoltage category ............................................................................................................................. II
RFI suppression ...................................................................................... conforming to EN 55011 Class B
Noise immunity ..................................................................................... conforming to DIN EN 61326/A1
Mains voltage Axio Lab.A1 ....................................................................................... 100 to 240 V ±10 %
..........................................................................................
Conversion of mains voltage unnecessary!
Mains frequency ........................................................................................................................ 50/60 Hz
Power consumption Axio Lab.A1 .................................................................................................. 110 VA
Fuses to IEC 127
Microscope stand Axio Lab.A1 ......................................................................... 2 x T 3.15 A/H, 5x20 mm