START-UP
Axio Lab.A1
Installing standard components
Carl Zeiss
04/2013 430037-7144-001
57
3.1.9
Attaching condenser
x
Move the stage carrier with focusing drive to
the higher stop position.
CAUTION
The objectives should not collide with
other parts.
x
Swivel out front lens (if shiftable) on condenser
using lever (Fig. 3-15/
7
).
x
Remove both centering screws (Fig. 3-15/
5
) on
the condenser carrier until their ends are no
longer visible.
x
Push the condenser carrier (Fig. 3-15/
3
) with
drive knob for vertical adjustment (Fig. 3-15/
2
)
down as far as it will go.
If using an overview fixture, make sure that this
does not come to rest on the luminous-field
diaphragm.
x
Insert the condenser (Fig. 3-15/
8
or
9
) between the condenser carrier (Fig. 3-15/
3
) and the stage
carrier (Fig. 3-15/
1
). In doing so, align the screwed stud bolt on the underside of the condenser with
the groove (Fig. 3-15/
6
).
x
Press the condenser with the dovetail ring against the mainspring (Fig. 3-15/
4
) of the condenser carrier
until the condenser sits horizontally on the condenser carrier.
x
Position the condenser on the carrier so that the screwed stud bolt sits against the groove (Fig. 3-15/
6
)
at the front.
x
Insert the centering screws until they engage with the dovetail ring of the condenser.
For mounting other types of condensers, proceed analogously.
Fig. 3-15
Attaching condenser