START-UP
Carl Zeiss
Mounting optional components
Axio Lab.A1
62 430037-7144-001
04/2013
3.2
Mounting optional components
The microscope must be switched off and unplugged before starting work.
Upon completion of work the respective modules must be restored to function (see Sections
3.1 to 3.4).
3.2.1
Mounting the light intensive co-observer unit
The light-intensive co-observer unit is mounted on the Axio Lab.A1 with a main observer and one or two
co-observers in accordance with the separate instructions for use for multi-discussion facilities (Order No.
425145-7144-001).
3.2.2
Mounting polarizer D or filter
holder
x
Lift the condenser carrier together with its drive
knob upwards as far as it will go.
x
If necessary, unscrew locking and holding pins
together with the overview fixture from the
condenser carrier.
x
Hold the polarizer or filter holder (Fig. 3-26/
4
)
parallel to the underside of the condenser
carrier (Fig. 3-26/
1
) and screw the holding pin
(Fig. 3-26/
2
) of polarizer (Fig. 3-26/
4
) with the
angled adjusting lever (Fig. 3-26/
3
) into the
front left threaded hole below the condenser
carrier as far as it will go.
x
Screw the locking pin (Fig.
3-26/
6
) with
adjusting lever (Fig. 3-26/
5
) as far as it will go
into the rear threaded hole of the condenser
carrier.
The other components shown in the system
overview at this point must be mounted
analogously.
Fig. 3-26
Mounting the polarizer D