OPERATION IN EXTREME CONDITIONS
The operating range for the SE400 is 0º to +55ºC, 0 to 90% relative humidity.
The SE400 will operate in temperatures below 0°C (32°F), but the following temperature
effects must be kept in mind when using in such conditions:
There is potential for icing to affect performance. Take care to ensure the SE400 is
functioning properly during use. If abnormal operation – such as lack of response to
changes of breathing pattern – is observed, the user should leave the contaminated
area immediately. Any malfunctions should be investigated before returning the
respirator to service.
The capacity of the battery can be reduced significantly when used at temperatures
below 0°C (32°F).
S.E.A. does not guarantee the performance of the SE400 when used at temperatures
below 0°C (32°F).
SE-DATA
SE-Data is a comprehensive personal protection management tool, designed to help
organisations develop and implement a company-wide personal protections program. It
includes an extensive database of hazardous materials.
IMPORTANT WARRANTY INFORMATION
The warranty on the SE400 respirator is NULL AND VOID if:
The respirator is operated at any time without filtration media correctly attached
Filtration media are not replaced according to the manufacturer’s recommendation, an
established change out schedule or when prompted by the respirator’s alert functions
Any disassembly of the respirator other than as described in these user instructions is
carried out by any person other than a qualified S.E.A. technician.
IN CASE OF HEAVY IMPACT
If the respirator receives a heavy impact, such as being dropped:
Check that the filtration media are still firmly attached and tightened correctly
Perform a Self-Test to ensure the respirator functions properly
Inspect thoroughly for physical damage. If damaged, contact S.E.A. The unit should
also be leak tested.