not stable, we suggest that the user add a voltage regulator with matching power between the
mains and the instrument.
3. In order to ensure the treatment effect and normal service life of the instrument, please use
the specified parts provided or suggested by the original manufacturer.
4. The instrument should not be placed in damp places or near water sources, and should not be
directly exposed to sunlight.
5. Do not place the instrument near a strong heat source, as this may affect the life and normal
use of the instrument.
6. Before treatment, please remove all metal objects from the body first, so as to avoid
unexpected conditions and influence the curative effect.
7. Please do not use this machine on eyes, thyroid, parathyroid, testicles, abdomen of pregnant
women, and pacemakers.
8. Please turn off the power switch of the device, and ensure that the total power supply of the
device is turned off before the staff leaves after daily use, so as to ensure the safety of the
electrical products.
9. Clean the instrument with normal saline (do not use ethyl alcohol) after operation to ensure
its cleanliness and hygiene and prolong its service life.
10. Do not use this instrument if there is a prosthesis in your chest. It cannot be used if there is
metal stent in the body (excluding female’s IUD).
11. Women in pregnancy, lactation or menstruation can not use this machine.
12. People who are drunk can not use this machine.
13. Those who have cesarean section within half a year or those who have abortion within 3
months or natural delivery within 2 months can not use this machine.
14. If the suction is set at too high level accidentally during the treatment, you should stop the
treatment immediately and reset the energy level.
15. Set the energy at lowest level at the beginning and slowly raise it after the customer adapts.
16. The treatment area should be kept moisturized during the operation and do not perform the
treatment on dry skin. Please use filter cotton before using grease cup in case essential oil
causes damages to the machine.
17. Use machine or train machine operators in strict accordance with instructions in the user
manual.
5. Troubleshooting & Solutions
1. The instrument cannot be started, and the button light on the back of the machine does not
work?
A. Make sure the power cord is connected to the socket with normal power supply.
B. Whether the fuse on the back of the instrument is loose or blown.
2. There is weak or no drawing force when you use the vacuum cup or grease cup?
A. Please turn off the machine and check whether the dedicated filter should be replaced.
B. Please check the hose connecting cup to machine body, which may be broken.
C. Please check whether the container of the filter is screwed tight and rubber rings are worn
down. The case can occur if the area is not airtight.
D. If methods above can not solve such problem, please contact machine distributor for help.
3. No RF output from the instrument?
A. Please check whether the cord connects the handle to the machine properly
B. Please check whether the treatment area has been cleaned. Grease or essential oil products
may cause poor contact between the handle and human body to cause this problem.