Serial Communications
Page 7-23
_________________________________________________________________
continue from that point. If no sequence is active, the
command will be ignored.
@ TS , {AAA} [CHK]<CR>
If a sequence is already active, the sequence will advance
to the next point and the timed sequence will continue from
that point (if it is the last point, the sequence will be
terminated). If no sequence is active, the command will be
ignored.
Manual Sequence (MS)
This command causes a manual, operator stepped sequence to be
activated. The point selected in the {point} field of the
sequence identified by {seq name} will be activated and will
remain active until another point is activated, calibrations are
stopped or a timeout occurs.
If a sequence point is active when the Manual Sequence command
is sent, it will be terminated and superceded by the new
sequence and/or point. It is not necessary to use the Stop
command before sending a Start Point command. A more detailed
description of the usage of the {seq name} and {point} fields
can be found on page 25.
Command Format
Command
@ MS , {AAA} , {seq name} , {point}, [CHK]<CR>
Where:
{seq name} =
The name of the sequence.
{point}
=
The point number.
Response
Good:
<ACK>
Error:
<NAK>[ER][<CR>]
Comments:
Leaving the {seq name} and/or {point} blank will have the
following results:
@ MS , {AAA} , {seq name} , [CHK]<CR>
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......