Internal UV Ozone Photometer
Page 9-6
After determining what mode of operation is to be used, the mode
should be selected by clicking the appropriate checkboxes in the
Photometer Parameters screen, as described on Page 12. Other
than customizing the analog outputs for a specific application,
the operating mode is the only setup function that is normally
necessary to operate the 2010
D
's Photometer.
Unless Continuous Mode is selected (in which case the Photometer
is always active), the
2010D
will automatically activate the
Photometer any time that an ozone point or sequence begins. The
Photometer
will
not
be
activated,
however,
during GPT
calibrations.
The ozone value, as measured by the Photometer, will be
transmitted from the three analog outputs and will also be
presented as the "Measured Ozone" value in the Photometer section
of the Ozone Calibration Status screen (see Figure 9-). This
section of the status screen (which may also be called up by
selecting "Photometer Status from the "Status" menu) not only
presents the current ozone concentration, but also provides the
current readings of the various sensors and solenoid valves
associated with the Photometer.
Figure 9-3 The Ozone Status Screen
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......