Internal UV Ozone Photometer
Page 9-12
Figure 9-6 Photometer Parameters, Page 1 of 3
In the upper left corner of the Photometer Parameters screen, the
operating mode of the Photometer can be changed by clicking the
following checkboxes:
Ozone Servo Control
When this box is checked, the ozone
measured by the Photometer will be used
to automatically adjust the 2010
D
's
ozone
generator
while
an
ozone
calibration is being performed.
Continuous Monitor
By checking this box, the Photometer
will remain active continuously while
power is applied to the
2010D
.
Temp. / Pressure Correction
The Photometer is normally operated with
temperature/pressure correction enabled,
however under certain circumstances,
such as when comparing the 2010
D
's
Photometer
with
an
external
ozone
monitor
that
does
not
support
temperature/pressure correction, it may
be desirable to temporarily disable this
feature by unchecking this box.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......