Installation
Page 4-10
Source gases are connected to Source Gas Input Ports, 1 through
4 and optionally Source Gas Input Ports 5 and 6.
Source gas
pressures must be regulated to 20 - 25 psig to prevent damage to
the Model
2010D
.
Cap all unused Source Gas Input Ports.
An optional Sabio Instruments Model
2010D
or Model 2500
Permeation Oven may be connected to the optional Perm Tube Input
Port.
The Purge port should be connected to the system exhaust
manifold. Instrument exhaust ports should also be connected to
the system exhaust manifold.
On all versions of the Model
2010D
calibrators, zero gas is
available through the Span Gas Out ports when a zero point is
produced.
The Vent Output Port should be connected to the system exhaust
manifold. Instrument Span Gas Input ports are connected to the
remaining Span Gas Out Ports, 1 through 7 starting with port 1.
Cap all Span Gas Out ports that do not have connections.
Power
The standard operating voltage for the Model
2010D
calibrators is
96 - 264 VAC at a frequency of 50/60 Hz. The power consumption
for the Model
2010D
is approximately 150 - 300 VA. Ensure that
the power source is rated properly.
An adequate earth ground
must be available through the AC power plug to ensure safe
operation of the unit and to prevent electrical shock hazard.
Parallel Printer Data Port
A standard, off-the-shelf, parallel port printer data cable with
a DB-25 connector on the Model
2010D
printer port end and a
Centronics port connector on the other end will generally work
for connections between the Model
2010D
and the printer in most
instances.
The Parallel Printer connector J1 on the rear panel of the Model
2010D
is a female type DB-25 connector.
Summary of Contents for 2010D
Page 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Page 3: ......
Page 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Page 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Page 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Page 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Page 178: ......