46
a. Add a new image to the register
To add a new image to the patient register, click on the icon
. A search window will be opened.
Select the image you want to add to the register and click "Open".
b. Delete an image from the register
METRIC CALIBRATION
To delete an image, select the image you want and click
to delete it permanently.
c. Open an image from the register
To open an image from the register, select the image that you want to open and click on the icon.
The process of metric calibration must be performed at least one time after
installation.
The metric calibration depends on the radiographic technique used. The
metric technique must be performed every time the technique is changed.
The intraoral calibration is unique for any device and it depends on the radiographic technique
used. There are two basic periapical radiographic techniques: the Bisector Technique and Paralle
-
lism Technique
Parallelism Technique or Long cone
In this technique, the Saevo Slim is placed in parallel to the axis plan of the tooth with the help of
a positioner and the focal point of the x-ray generator must be positioned at a distance around
40cm of the sensor.
a. Radiographic techniques
This technique produces radiograph with minimum distortion, showing objects that are being x-rayed
in their true anatomic relation and size.
SYSTEM OPERATION