54
CLEANING AND DISINFECTION
Before starting the cleaning and disinfection procedure, disconnect the sensor
from the USB port of the computer to avoid damage to your equipment.
For your protection, during the cleaning and disinfection process, it´s man
-
datory the use of IPEs such as disposable gloves and safety glasses.
The cleaning and disinfection process must be performed at each patient change and before the
storage of the product.
When starting the process, check the existence of visible dirt, such as blood or saliva.
Clean carefully the Saevo Slim and the first 30cm of the USB cable.
For the cleaning, use a clean and soft cloth damp with neutral soap and then perform the drying
with a clean and soft cloth or with a paper towel.
For the disinfection process of the Saevo Slim, use disinfectants towels that have active compo
-
nents based on Didecyldimethylammonium chloride respecting the contact time indicated by the
manufacturer.
There's no cycle limits or time of application that the Saevo Slim and its parts can tolerate during
the cleaning and disinfection process, following the instructions of this manual.
Don't use sanitizing or cleaning agents as a spray.
Don't dip the Saevo Slim sensor in liquid sanitizing.
Don't sterilize the Saevo Slim sensor using an autoclave, thermal sterilizer.