O p e r a t i o n s a n d I n s t r u c t i o n M a n u a l
63
S A G E M E T E R I N G , I N C .
REV. 10-SIX/SRX
HART communications are optionally provided on
some versions of the Sage Prime Thermal Mass Flow
Meter. HART is designated by the code “HART” in
the model number.
HART
RIO DISPLAY
NOTE: HART will only be available if the 4-20 mA
output is externally powered. A minimum resistance
of 250 ohms is required between the power supply
and the flow meter.
Next to the Temperature units of measurement (C or
F) there are three pixel points.
See illustration above
.
• The top indicator is lit if HART is active.
• Middle indicator lit or flashes during active data
transfer.
• The bottom is lit when the instrument is under
HART control for manual settings or testing.
Lit when HART is being transmitted.
Lit when HART is being received.
F
HART Active
Data Transfer
Under HART control