- 33 -
Press the
SET
key to skip over a parameter to the next.
Absolute Pressure Calibration
In the calibration mode, the “ABS” symbol will display at the
pressure section, the absolute pressure value will flash. (The
default value is 0.00 inHg)
Press the
C
key or
MAX/MIN/-
key to increase or
decrease the absolute pressure value (in increments of 0.01
inHg).
Press and hold the
C
or
MAX/MIN/-
key for three
seconds to increase or decrease rapidly.
Press the
ALARM
key to reset current value.
Example
: The calibrated pressure sources measure 28.37
inHg. The display console pressure reads 28.75 inHg.
Offset = 28.37 - 28.75 = - 0.38 inHg
Relative Pressure Calibration
In the calibration mode, press the
SET
key again, the “REL”
symbol will display at the pressure section, the relative
pressure value will flash. (The default value is 0.00 inHg)
Press the
C
key or
MAX/MIN/-
key to increase or
decrease the relative pressure value (in increments of 0.01
inHg).
Press and hold the
C
or
MAX/MIN/-
key for three
seconds to increase or decrease rapidly.