6. If the calibration was
not
successful, one of the error messages below will appear. Take
the indicated action to correct the problem, then perform a new calibration.
"
Err0
" - The calibration test weight or the adjusted keyed-in weight is larger than the full
capacity of the scale. Change the calibration test weight or check the input data.
"
Err1
" - The calibration test weight or the adjusted keyed-in weight is smaller than 1% of
the full capacity of the scale. Change the calibration test weight or check the input data.
"
Err2
" – The calibration test weight does not match the keyed-in weight. This could also
be due to an internal fault and/or damaged load cell. Or you could have simply forgotten
to place the test weight on the scale before pressing the NET/GROSS key.