Technical Specifications 

 

 

Scale Indicator 

Input signal range:  

 

0mV 

 +30mV 

Sensitivity: 

 

  >0.2uV/grad 

Internal Resolution:  

 

Approximately 520,000 counts 

Display Resolution:  

 

Can be selected between 500-100,000 

System Linearity:  

 

Within 0.01% of FS 

Loadcell excitation Voltage:  

+5 V

DC

 (MAX current: 85mA) 

 

Loadcell 

Sensitivity:  

 

 

0.3mV/V --- 3mV/V (must be fit to >0.2uV/display grad.) 

Input Resistor:    

 

60 

 

Output Resistor:  

 

<10 K

 

 

Temperature 

Operation: 

   

5

o

C - 35

 o

Storage:    

10

o

C - 70

 o

Humidity:    

95%RH (no condensation) 

 

Power  

 

Battery:  

 

 

6V4AH lead acid battery, 30 hrs continuous use 

When the voltage of battery is below 5.7V, the “Low Bat” annunciator will be lighted, plug in AC adapter to charge 
the battery.  When “Lo.bAt” and actual weight is displayed alternately, this indicates the voltage of battery is below 
5.5V and the scale will be turned off in two minutes automatically.  
 
AC 

Adapter: 

   

10.5-12V

DC

 600mA, with central positive, 5.7 VDC -9 VDC 

 Error Messages 

 

Error Message  

Definition 

Required 
Solution/Troubleshooting 

 
0¯ ¯ ¯ ¯: 

Weight above range for calibrated 
zero point. 

Remove load before zeroing 
Or 
Recalibrate the scale. 

 
 
0_____: 

Weight below range for calibrated 
zero point. 

Remove load before zeroing 
Or 
Recalibrate the scale. 

 
_____: 

Indicates an under range condition  Remove all loads and zero the 

scale. 

 
¯ ¯ ¯ ¯: 

Capacity exceeded 

Remove the load and try again.  If 
the load is great than 6 kg a scale 
with a larger capacity is required. 

CAL-Er: 

Calibration error 

Restart calibration 

Lo.bAt: 

Low Battery 

Recharge the battery. Upon initial 
use, it is recommended to charge 
battery for 8 hours prior to use. 

EEP.E0 

EEPROM can’t be accessed 

Replace S100 Indicator 

EEP.E1 

Configuration settings  have 
changed and not been stored 

Calibrate the scale to store 
settings 

EEP.E2 

P7-P9 settings exceed scale’s 
normal range 

Reconfigure P7-P9 setting  

Summary of Contents for SBI-100

Page 1: ...Model SBI 100 Operation Manual Manual xxxxxx Issue xx...

Page 2: ...DOUBT SEEK ADVICE FROM A QUALIFIED ELECTRICIAN Routine maintenance To avoid the possibility of electric shock or damage to the machine always switch off the machine and isolate from the power supply...

Page 3: ...pecifications 15 Error Messages 15 Declarations of Compliance 16 Contact Information Technical Support 16 Product Identification Table Part Number Capacity Region Power Supply s 816965001620 75 kg x 0...

Page 4: ...t Power Adapter Socket Power Supply Connecting the Indicator to the Platform 6VDC 500mA 230V 50Hz Wire the cable attached to the base as shown if necessary 4 2 3 1 1 RED 2 RED 3 RED 4 RED 1 Red Excita...

Page 5: ...0 Weight is zero On Off Zero On Off Zero kg lb Unit of measure Flashing LED means that the weight reading is not stable Units kg lb Unit of measurement NET Tare is activated Hold Print Hold Print CH B...

Page 6: ...tion Turning On and Zeroing the Scale On Off Zero ZERO 0 Turning off the scale On Off Zero Selecting Unit of Measure Units kg lb kg lb Using the Tare 1 2 Tare NET 3 NET 4 NET Removing the Tare On Off...

Page 7: ...old function with manual release This function needs to be setup in Parameter P3 1 shown on page 10 1 Hold Print Hold 2 3 Removing Hold Hold Print Hold Print Function For communications to a printer o...

Page 8: ...s Hold Print Entering setup On Off Zero Units Selecting parameter Tare Changing data within the parameter Units Saving data Tare Exiting setup On Off Zero ZERO 0 kg Moves flashing digit Units Change f...

Page 9: ...ctable hold window from 3 to 50 divisions Will hold display reading once stable within a selectable weight range to release the hold button must be represses P3 0 P3 1 P3 2 P3 3 to 50 P4 x RS232 Seria...

Page 10: ...tare LF H1H2H3 CR ETX U CR Changes the units LF u1u2 CR LF H1H2H3 CR ETX L CR Activates the hold function LF H1H2H3 CR ETX X CR Switches off the scale Indicator switches off Unrecognised command LF CR...

Page 11: ...mple Gross xxxxx0 18lb Tare xxxxxx0 18lb Net xxxxxxx0 18lb Format LF Gross reading minus decimal point unit CR EXT LF Tare reading decimal point unit CR EXT LF Net reading minus decimal point unit CR...

Page 12: ...olution 00 500 08 2400 15 7000 01 600 09 2500 16 7500 02 750 10 3000 17 8000 03 800 11 3500 18 10 000 04 1000 12 4000 19 12 000 05 1200 13 5000 20 15 000 06 1500 14 6000 21 31 N A 07 2000 P8 0 1 2 See...

Page 13: ...0 kg lb 5 Single point calibration enter the same weight in again and move to number 7 For 2 point calibration enter in the second calibration weight between 25 100 full capacity Tare 6 7 Tare 8 9 Tar...

Page 14: ...and LB P12 0 7 7 Power On zero range full scale 0 1 1 2 2 5 3 10 4 20 5 50 6 100 7 no limitation P13 00 15 3 Zero button range Full scale 00 1 06 20 12 5 01 2 07 no limit 13 10 02 3 08 1 14 20 03 4 09...

Page 15: ...ill be turned off in two minutes automatically AC Adapter 10 5 12VDC 600mA with central positive 5 7 VDC 9 VDC Error Messages Error Message Definition Required Solution Troubleshooting 0 Weight above...

Page 16: ...ications Le pr sent appareil num rique n met pas de bruits radio lectriques d passant les limites applicables aux appareils num riques de la Classe A prescrites dans le R glement sur le brouillage rad...

Reviews: