Technical Specifications
Scale Indicator
Input signal range:
0mV
~
+30mV
Sensitivity:
>0.2uV/grad
Internal Resolution:
Approximately 520,000 counts
Display Resolution:
Can be selected between 500-100,000
System Linearity:
Within 0.01% of FS
Loadcell excitation Voltage:
+5 V
DC
(MAX current: 85mA)
Loadcell
Sensitivity:
0.3mV/V --- 3mV/V (must be fit to >0.2uV/display grad.)
Input Resistor:
≥
60
Ω
Output Resistor:
<10 K
Ω
Temperature
Operation:
5
o
C - 35
o
C
Storage:
10
o
C - 70
o
C
Humidity:
≤
95%RH (no condensation)
Power
Battery:
6V4AH lead acid battery, 30 hrs continuous use
When the voltage of battery is below 5.7V, the “Low Bat” annunciator will be lighted, plug in AC adapter to charge
the battery. When “Lo.bAt” and actual weight is displayed alternately, this indicates the voltage of battery is below
5.5V and the scale will be turned off in two minutes automatically.
AC
Adapter:
10.5-12V
DC
600mA, with central positive, 5.7 VDC -9 VDC
Error Messages
Error Message
Definition
Required
Solution/Troubleshooting
0¯ ¯ ¯ ¯:
Weight above range for calibrated
zero point.
Remove load before zeroing
Or
Recalibrate the scale.
0_____:
Weight below range for calibrated
zero point.
Remove load before zeroing
Or
Recalibrate the scale.
_____:
Indicates an under range condition Remove all loads and zero the
scale.
¯ ¯ ¯ ¯:
Capacity exceeded
Remove the load and try again. If
the load is great than 6 kg a scale
with a larger capacity is required.
CAL-Er:
Calibration error
Restart calibration
Lo.bAt:
Low Battery
Recharge the battery. Upon initial
use, it is recommended to charge
battery for 8 hours prior to use.
EEP.E0
EEPROM can’t be accessed
Replace S100 Indicator
EEP.E1
Configuration settings have
changed and not been stored
Calibrate the scale to store
settings
EEP.E2
P7-P9 settings exceed scale’s
normal range
Reconfigure P7-P9 setting