background image

BiNAPS

 S

ensor module, Model 5500

Sensor cannula sample pack

 

(1 adult oral - nasal cannula  #5001 (with 

Filter

), 

1 adult nasal cannula

 

#5011 

(

with Filter), 1 adult divided nasal 

cannula #5018, 1 pediatric

 

oral - nasal cannula #5023)

User's Manual, Part No. 121321

BiNAPS and diagnostic cannula brochures

This product is for diagnostic purposes only and is not to be used as an

 

apnea

 

monitor or in a life supporting or life sustaining situation.

This

 

device is not intended for pediatrics and infants below two years of age for

 

the purpose of respiration or  SIDS monitoring. 

WARNING

Do not use this device with non-Salter Labs 

Diagnostic 

Cannulas.  The

 

use of  non-

Salter Labs 

cannulas  will  void  the warranty.

The patient should wear loose 

f

itting clothing, 

to minimize 

the chance of 

the device contacting the

 

skin.

 

GENERAL     INFORMATION

Airflow Ou

tp

ut

:

 Output is a nasal pressure airflow waveform.  Output is

 

  

connected to a polygraph's AC jackbox with 1.5mm safety connectors.  Snoring

 

can also be superimposed onto the airflow waveform as shown in the example

 

  

on the next page.  This is accomplished by setting higher values for the high

 

  

frequency filter and the sampling rate for the airflow channel.

 Snore Ou

tp

ut

:

 Output is a snoring waveform derived from snore vibrations on

 

  

the nasal pressure signal.  The channel has internal low frequency filter to

 

  

remove the airflow signal and to provide a flat baseline between snores.  Output

 

is connected to a polygraph's AC jackbox with 1.5mm safety connectors.

Hi /Lo Switches

:

 Are used as needed to optimize the airflow and snore signal

 

levels into a polygraph AC jackbox.  The "Hi" position will boost signal ampli-
tudes by a factor of  four, but will not boost signal amplitudes to DC input levels.

Plug each lead into 

the 

side port to verify signal. If green light fails to 

illuminate

,

 

 assume 

the device is 

defective and replace 

the 

BiNAPS

 device

A

fter

 testing, remove the leads from the indicator port 

INSTALLATION      INSTRUCTIONS

Nasal Airflow Pressure and Snore Transducer

 

Nasal Airflow Pressure and Snore Transducer

®

®

 

®

  

The Salter Labs BiNAPS Airflow Pressure Transducer is an accessory  intended

 

for use with polysomnography equipment during sleep disorder studies for the

 

purpose of detecting and amplifying breathing signals and detection of snoring

 

of 

a sleeping patient though a Salter Labs nasal cannula.

 

This device is intended to be set-up and maintained by credentialed sleep lab 

technicians or qualified medical personnel.

Clinical Benefit:

 The BiNAPS detects aiflow and snoring via a nasal pressure to 

monitor patients breathing patterns

PACKAGE CONTENT

CAUTION

INTEGRAL   CIRCUIT  TESTER  /  INDICATOR

INTENDED USE

1. Use Hook and Loop patches to secure the BiNAPS to the AC jackbox or

other convenient location.

2. Connect the "Airflow" leads to the selected polygraph AC jackbox inputs.

3. Connect the "Snore" leads to the selected polygraph AC jackbox inputs.

4. Apply the airflow sensor cannula to the patient. Insert the prongs into the

nares and loop the headset tubing over the ears and under the chin.  Slide

the bolo up under the chin for a comfortable fit. If necessary, secure the

headset tubing in position, e.g. Salter Labs Tender Grip #1005.

5. It is extremely important that the cannula sensor tips (prongs) do not

become blocked during installation or  recording.

6. Plug the sensor cannula safety filter into the input of the BiNAPS sensor

module with a slight twist to ensure a secure connection.

7. Adjust the polygraph settings per the table below. These settings are

recommended starting points. Polygraph and patient variables can

significantly influence the settings.

Polygraph & BiNAPS

             Settings

 High Frequency Filter

  Low Frequency Filter

 

(Time Constant)

     Sampling Rate

            Sensitivity

  BiNAPS Hi /Lo Switch

           Snoring

  70Hz or higher

       10Hz or lower

(0.016 sec. or  longer)

       70Hz or higher
           50µV/mm

       Adjust for optimal

   signal amplitude

        Airflow

 5Hz or higher

       0.05Hz or lower

 

(3 sec.  or  longer)

        10Hz  or  higher
              50µV/mm

   Adjust for optimal

 

signal amplitude

 Snore Output

Airflow output, filtered to allow snoring

Airflow output, filtered to eliminate snoring

Input  from  Filtered  Airflow

Sensor Cannula

  Polygraph

AC Jackbox

Reviews: