Marker menu
Marker
1 2 3 4 5 6
Marker
Normal
Delta
Noise Marker
Off
On
Marker- >
Off
All Off
>
· [Marker 1 2 3 4 5 6]: Press [Marker]→ [Marker 1 2 3 4 5 6] to change
various markers. The selected marker will be underlined.
· [Normal]: Press [Marker]→[Normal] to enable the normal mode of the
current marker.
· [Delta]: Press [Marker]→[Delta] to enable the delta mode of the current
marker. In this case, the frequency difference and amplitude difference
(time difference under zero span) between the differential marker and
reference marker will be displayed. The amplitude will be displayed in dB.
· [Noise Marker Off On]: Press [Marker]→[Noise Marker Off On]. The
noise marker will indicate the noise power normalized to 1Hz bandwidth
near the active marker. In this case, the “Sample” mode of the detector will
be enabled. If the noise marker is enabled, the unit of the marker reading
will automatically change into dB/Hz.
· [Marker→]: Press [Marker]→ [Marker→] to open the marker function
menu. By using such marker functions, the user can change the instrument
display, with the marker as the reference. Refer to [Marker→] menu for
details.
· [Marker off]: Press [Marker]→ [Off] to disable the current active marker.
· [All Off]: Press [Marker]→ [All Off] to disable all active markers.
[Special note]: Marker functions are not available in the RSSI mode.