Table 6.1 List of instrument feature error information
Error code
Error Description
1052
Data error Only for digital-based band interface: Digital I/Q input data error.
1053
FIFO overload Only for digital-based band interface: High external input sampling
rate.
1054
Input overload Signal power of RF input port is out of specified range.
1055
No reference. The signal generator does not detect any required external reference
single input.
1056
OVEN OCXO is below working temperature, and a few minutes later after power on,
OCXO enters normal working state, and the error information disappears.
2) Type of error information
Error event is only provided with an error information, and types of error information are as follows:
Query error (–499 to –400): Instrument output queue control detects the message exchange
protocol error described in IEEE 488.2, Chapter 6. Now the event state register‘s query error bit (bit2)
is set (refer to IEEE 488.2, 6.5 for details). Now data can not be read successfully from output queue.
Instrument feature error (399 to –300, 201 to 703, and 800 to 810): Instrument operation fails,
which may be caused by abnormal hardware/firmware state, and this error code is often used in
instrument self test. Now the event state register‘s instrument feature error bit (bit3) isset.
Execution error (-299 to –200): Fault is detected during configuration of the instrument. Now the
event state register‘s execution error bit (bit4) is set.
Command error (-199 to –100): The instrument detects syntactic error during command parsing,
which is caused by incorrect command format. Now the event state register‘s command error bit (bit5)
is set.
6.4 Repair methods
6.4.1 Contact us
When the S1465 series signal generator is faulty, first view the error information and store it, find
possible causes and refer to the methods provided in the section “6.2 Fault diagnosis and
troubleshooting”, so as to troubleshoot the signal generator in advance. Please contact our service
center according to the contact information below in the case of fault unsolved, and provide collected
error information, so that we can help you to solve your problem as early as possible.
Contact information:
Tel:
+886.909 602 109
Website:
E-mail:
Address:
No.367 Fuxing N. Road, Taipei 105, Taiwan
6.4.2 Packaging and transport
If you have any intricate problem with your signal generator, contact us by phone or fax. After
necessary repair is confirmed, pack the signal generator with original packaging materials and put it
in a box before the following steps.
1) A detailed description of fault symptom should be provided and attached to the signal generator.
2) The signal generator should be packed by the original packaging materials, so as to minimize
damages.
3) Four corners of external packaging box should be cushioned before the instrument is put in the
box.