Parameter
1
X
2
Y
3
R
4
θ
5
Xh1
6
Yh1
7
Rh1
8
θh1
9
Xh2
10
Yh2
11
Rh2
12
θh2
13
Noise
14
A1
15
A2
16
A3
17
A4
18
Freq
Values are returned as ASCII floating point numbers with units of Volts, degrees or Hz.
This command is a query only command.
SNAP?i,j,{k,l,m
,n,o,p,q,r,s,t,u}
The SNAP? command records the values of no more than 13 parameters at a single instant.
For example, SNAP? is a way to query values of <X>, <Y>, <R>, <θ> and <F> which are
taken at the same time. This is important when the time constant is very short. Using the
OUTP? or OUTR? commands will result in time delays, which may be greater than the
time constant, between reading two different parameters.
The SNAP? Command requires at least two parameters and at most thirteen parameters.
The parameters i, j, k, l, m, n, o, p, q, r, s, t, u select the parameters below.
i,j,k,l,m,n,o
,p,q,r,s,t,u
Parameter
1
X
2
Y
3
R
4
θ
5
Frequency
6
Xh1
7
Yh1
8
Rh1