Fig.63 The Configuration Area of Internal Frequency Sweep
Table 5. Internal Sweep frequency Reference Configuration
Int. Sweep Type
Linear
Log
Int. Sweep Start
(Hz)
Setting the start frequency of sweep manually. The frequency range is
from 1mHz to 102kHz and the minimum resolution is
1mHz.
Int. Sweep Stop
(Hz)
Setting the stop frequency of sweep manually. The frequency range is
from 1mHz to 102kHz and the minimum resolution is
1Hz.
Linear Step Time
(ms)
Setting the step time for every sweep manually. The step range
is from 1ms to 100s and the minimum resolution is 1ms.
Int. Sweep Step
(Hz)
Setting the step frequency for every sweep manually. the step
range is from 1Hz to 102kHz and the minimum resolution is 1Hz.
Int. Sweep Run
Mode
Single
Loop
Stop
Note that:
After finishing all the settings, click the button
to finish the configuration for input signal.All
settings can be changed at the same time and you just need to click this button once.
5. Dynamic Reserve and Sensitivity Configuration