6.4 Amplitude Accuracy and Flatness
This test measures the amplitude accuracy and frequency response.
Device
Use the function signal generator to provide an accurate frequency and a sine wave.
Use one cable with BNC connector to connect the output of the function signal generator to the A input of the
lock-in. Use another cable with BNC connector to connect the reference output of the function signal generator to
the REF IN of the lock-in.
Set the function signal generator to:
Function
Sine
Frequency
1 kHz
Amplitude
1 Vrms
Offset
0 V
Output
High Z
Sweep
off
Modulation
none
Steps
1. Turn the lock-in off and on to restart it.
2. Set <Filter> to 24 dB/oct.
3. Amplitude accuracy is verified at 1kHz and various sensitivities. Set the sensitivity of the lock-in and the
amplitude of the function signal generator in the following sequence:
Sensitivity
Amplitude
1 V
1.0000 Vrms
200 mV
200.00 mVrms
100 mV
100.00 mVrms
20 mV
20.000 mVrms
10 mV
10.000 mVrms
a) Set the function signal generator to the amplitude shown above.
b) Set <Sensitivity> of the lock-in.
c) Wait 10 seconds, then record <R>, then test the next group of the sensitivity-amplitude.
d) Repeat the steps from 3)-a) to 3)-c) until the Sensitivity-Amplitude test is completed.
4. Frequency response is checked at frequencies above 1kHz. Set the frequency in the following sequence:
Test Frequency
24 kHz
48 kHz
72 kHz