Tel: 886. 909 602 109 Email: sales@salukitec.com
www.salukitec.com
104
-50.00
-100.00
c) Waiting for 10 seconds and record the reading of the DVM. Then, test next group data.
d) Repeat steps 3-b) to 4-c) until finishing the tests for CH1. Then, connect the DVM to CH2 and continue to
finish the tests for CH2.
4. Follow the steps below:
a) use signal cable to connect the DVM to the Lock-in AUX_OUT 1 on rear panel.
b) set <DA Output> of <AUX Output> in the following sequence
DA Output (V)
10.000
5.000
0.000
-5.000
-10.000
d) Wait 10 seconds and record the reading of the DVM. Then, test next group data.
e) Repeat steps 3b to 3c until finishing the tests for DA1. Then connect the DVM to DA2, DA3, DA4 to
finish the tests for DA2, DA3, DA4.
5. Follow the steps below:
a) Set <Monitor> in <Display> to <Input>.
b) use signal cable to connect the AC regulator OUT to AUX_IN 1.
c) use signal cable to connect the AC regulator OUT to AUX_IN 1.
Voltage (V)
10.000
5.000
0.000
-5.000
-10.000
d) Wait 10 seconds and recode the reading of <A1> at the top of the screen. Then, test the next group data.
e) Repeat steps 4c to 4d until finishing the tests for A1. Then connect the AC Regulator OUT to A2, A3, A4 and
finish the tests for A2, A3, A4.
6. End the tests for DC inputs and outputs and fill the results in the performance test record form.
6.9 Input Noise
This test measures the input noise of the lock-in.