88
Fig.73 The Harmonic Waveform Display
5.3 Usage Examples
This example will show how to use SE1022 PC software to configure parameters, and observe and record the value
of R/X/Y/
θ
.
First, connect the SE1022 and PC according to the software usage guidance in 5.2 and then you can start to
configure.
Suppose that the lock-in amplifier configurations are shown in Table 14.
Table 14. Configuration Instance
Input Source
Single-ended voltage input
Input Amplitude
40mV
Input Coupling
AC
Input Notch Filter
Close
Dynamic Reserve
Low Noise
Sensitivity
Auto Sensitivity
Reference Source
External reference, 1000Hz
Reference Signal Trigger Way
TTL rising edge trigger
Phase Shift Angle
0
Time Constant of Low Pass Filter
300ms
Roll Off
12 dB/oct
Sampling Rate
125Hz