99
<Ref. Source>
Set to <Internal>
<Ref. Frequency>
Set to 1Hz
<Sensitivity>
Set to <1mV> through the spinning knob
3. Wait 10 seconds, then record the value of R.
4. Change the setting:
<Coupling>
Set to <DC>
5. Wait 10 seconds, then record the value of R.
6. End the test and fill the result into the performance test form.
6.3 Common Mode Rejection
This test measures the common mode rejection of the lock-in.
Device
Use the internal oscillator sine output to provide the signal.
Connect the [SINE OUT] to both A and B inputs of the lock-in.
Use equal length cables from A and B to a BNC TEE.
Connect the cable from the [SINE OUT] to the TEE. Do not use any termination.
Steps
1. Turn the lock-in off and on to restart it.
2. Press the buttons in the following sequence:
<Sensitivity>
Set to <1 V> through the spinning knob
<Ref. Source>
Set to <Internal>
<Ref. Frequency>Set to 100 Hz
3. Wait until <R> is steady and <R> should be 1Vrms (within 3% error)
4. Press the buttons in the following sequence:
<Coupling>
Set to <DC>
<Source>
Set to <A-B>
<Sensitivity>
Set to <200
μ
V> through the spinning knob
5. Wait 10 seconds, and then record the value of R.
6. End the test. The common mode rejection equals to 20lg(1/R), where R is in Volts(V). Fill the results in the
performance test form.