3.3 [GAIN/TC]
Press [GAIN/TC] button, then enter the submenu as is shown in Fig.18. There are 5 function settings in the
submenu, including <Sens>, <DR>, <TC>, <Filter > and <Sync>.
Fig.18 [GAIN/TC] Submenu
3.3.1 < Sens>: Setting the Full Scale Sensitivity
Choose an appropriate full-scale sensitivity according to the amplitude of the input signal. Adjust the value through
the soft keys or spin knob.
Table 1. Full Scale Sensitivity
2 nV/fA
500 nV/fA
100 μV/pA
20 mV/nA
5 nV/fA
1 μV/pA
200 μV/pA
50 mV/nA
10 nV/fA
2 μV/pA
500 μV/pA
100 mV/nA
20 nV/fA
5 μV/pA
1 mV/nA
200 mV/nA
50 nV/fA
10 μV/pA
2 mV/nA
500 mV/nA
100 nV/fA
20 μV/pA
5 mV/nA
1 V/μA
200 nV/fA
50 μV/pA
10 mV/nA
The charge of <Sensitivity> will change the dynamic range of the system. Besides, it will impact the output of
CH1&CH2. If the measure signal is current, the full scale acts as the corresponding current unit, <100mV/nA> in
default.
3.3.2 < DR>: Setting the Dynamic Reserve Mode
<Low>
Low dynamic reserve mode
<High>
High dynamic reserve mode.
<Normal>
Normal dynamic reserve mode. Generally, this mode is recommended.