53
Channel 2
Choose CH2 to output the required values. The value types contain the
X/Y/R/θ value of signal.
Offset (%)
The range is -100% to +100%, the minimal step is 0.01%, 0.00% in default.
Expand
The range is 1 to 256, 1 in default.
9. Sine Out Configuration
The sine out configuration is in the red box in the Fig.46.
The amplitude of the sine out can be set manually from 100 mVrms to 1.000 Vrms. The minimum accuracy is 1 mVrms.
Fig.46 The Configuration Area of Sine Out Channel
10. Data Storage
The software has the function of data storage. Users can choose to save the data or not that SE1201 collects in a period.
The saved data contains the value of X, Y, R,
θ
and Freq.
The steps for choosing to save data or not:
1. The data is saved in Excel. The file name is “Data_recorded_excel.xls”. It is saved in the program directory.
2. When the software is executing, click the button
in the red box in Fig.47. After clicking this button
and it displays
. The software is saving the current collected data.