Reference Source
External reference, 1000Hz
Reference Signal Trigger Way
TTL rising edge trigger
Phase Shift Angle
0
Time Constant of Low Pass Filter
300ms
Filter Slope
12 dB/oct
Sampling Rate
10Hz
The step to finish the configuration in table 8 are as follows:
1. According to Table 8, choose the following options in
“
Input Setting
”
, including
“
Input Source
”
,
“
Input
Coupling
”
and
“
Input Notch Filter
”
with other options in default, as is shown in Fig.48. At last, click the button
to finish the configurations.
Fig.48 Input Configuration
2. Choose the following options in
“
Reference Configuration
”
, including
“
Phase
”
,
“
Reference Source
”
,
“
External
Ref Trigger
”
with other options in default, as is shown in Fig.49. At last, click the button
to finish