54
3. Click the button
again, and the button status will switch back to
again. This means the
software stops saving collected data.
4. In box
, you can modify the sampling rate of the current display and data storage. The input range is
0.1s - 100s.
Fig.47 The Configuration Area of Data Storage
5.3 Usage Examples
This example will show how to use SE1201 PC software to configure parameters, and observe and record the value
of R/X/Y/
θ
.
First, connect the SE1201 and PC according to the software usage guidance in 5.2 and then you can start to
configure.
Suppose that the lock-in amplifier configurations are shown in Table 8.
Table 8. Case Configuration
Input Source
Single-ended voltage input
Input Amplitude
40mV
Input Coupling
AC
Input Notch Filter
None
Dynamic Reserve
Normal
Sensitivity
50mV/nA