39
b. Splicing the fiber quickly.
c. The electrode position is accurately measured 16 times of electrode
stabilize.
8.4 Electrode setting
The splicing loss will be enlarged and splicing strength will be reduced when the
discharging times exceed the electrode life. The electrode is worn by use and must
be regularly cleared according to the concentration of the oxide. Set a reminder
when the electrode was used for 2000 times, and it is recommended to update new
electrode bar when splicing over 2000 times.
When over 3000 times, there will be [please change the electrode bar] reminder
when turning on.
When change the electrode bar, please press [replace electrode] in [electrode
setting] or turn off the power and change.
Lose the screw on the electrode, take off the old electrode bars.