For example: WrtCmd (“CORR:SPOT1:FREQ 2KHZ”) Set the frequency of
correction spot 1 as 2KHZ.
NOTE: <value> should be ranged from 20HZ to 1MHZ (
STB8827
C),
or return format will report errors.
103
Query syntax: CORRection:SPOT<n>:FREQuency?
Return format: <NR3><NL^END>
The CORRection:SPOT<n>:OPEN command is execute the open correction for
the current correction spot.
Command syntax: CORRection:SPOT<n>:OPEN
Where,
<n>:
One of the 201 correction spots.
For example: WrtCmd (“CORR:SPOT1:OPEN”) Execute open correction for the
current correction spot 1.
The CORRection:SPOT<n>:SHORt command is used to execute short correction
for the current correction spot.
Command syntax: CORRection:SPOT<n>:SHORt
Where,
<n>:
One of the 201 correction spots.
For example: WrtCmd (“CORR:SPOT1:SHOR”) Execute short correction for the
current correction spot 1.
The CORRection:SPOT<n>:LOAD:STANdard command is used to set the
standard reference of the current correction spot. The
CORRection:SPOT<n>:LOAD:STANdard query returns the current standard
reference of the current correction spot.
Command syntax: CORRection:SPOT<n>:LOAD:STANdard <REF. A><REF. B>
Where,
<n>:
One of the 201 correction spots.
<REF. A> can be NR1, NR2 or NR3 data format and taken as the standard
reference of the primary parameter.
<REF. B> can be NR1, NR2 or NR3 data format and taken as the standard