DC bias voltage source
0V— ± 5V Minimum resolution: 0.5mV, Accuracy: 1% x preset 5mV
0mA—± 50mA Minimum resolution: 5
μ
A
6.3 Measurement accuracy
Test accuracy includes stability
、
temperature coefficient
、
linear degree
、
test
repeatability and calibration inter-error.
Check the accuracy of instrument should be under the following circumstances
:
a
.
b
.
c
.
d
.
e
.
warm-up time
:
≥
30 min
cable
:
0m, 1m
correct open and short correction after warming up
DC bias is in the position of “OFF”
The range works in “AUTO” to select correct test range
6.3.1 Accuracy of
│
Z
│
,
│
Y
│
, L, C, R, X, G, B
The accuracy Ae of
│
Z
│
,
│
Y
│
,
L
,
C
,
R
,
X
,
G and B are expressed as
:
A
e
= ± [A
L
× A+
(
K
a
+K
b
+K
c
)
×100+ K
d
+K
f
]×K
e
[%]
A
:
basic test accuracy
(
figure A
)
A
L
: level correction factor
(
table A
)
K
a
:
impedance rate factor
(
table B
)
K
b
:
impedance rate factor
(
table B
)
K
c
:
calibrated interpolating factor
(
table E
)
K
d
:
cable length factor
K
e
:
temperature factor
(
table G
)
K
f
:
scan fixture modification factor
(
no adding
:
K
f
= 0, adding
:
K
f
= 0.2
)
Using condition of L, C, X, B accuracy A
e
:
D
x
(
test value of D
)
≤
0.1
Using condition of R, G accuracy A
e
:
Q
x
(
test value of Q
)
≤
0.1
When D
x
≥
0.1, accuracy factor A
e
of L, C, X, B should be multiplied by
1
D
x
2
When Q
x
≥
0.1, accuracy factor A
e
of R, G should be multiplied by
1
Q
x
2
Accuracy of G can only be available in G-B combination.
6.3.2 Accuracy of D
The accuracy of D is given by the formula below
:
A
100
e
D
e
=
The formula is only available when D
x
≤
0.1.