STB8827
Series Operation Manual Chapter 7 Command Reference
1k
77
Ω
10k
Ω
100k
Ω
100
μ
4 Standard
inductor
H
0.02%
1mH
10mH
100mH
(
0
~
1000
)
5 Frequency
counter
MHz
0.5
6 Digital
Multimeter
%
7
Insulation resistance meter
500V 10 levels
0.25kW
(
0
~
500
)
8 Hipot
Tester
V
6.6.3 Function check
Ensure function keys, displayer and terminal etc. can work normally.
6.6.4 Test signal level
Adjust multimeter in AC voltage range, where one test cable is connected to H
CUR
and the other is connected to ground terminal. Change level as: 10mV, 20mV,
100mV, 200mV, 1V, 2V, the reading should meet the demand of test signal level in
this chapter.
6.6.5 Frequency
Connect frequency meter to ground terminal. The test terminal of the frequency
meter is connected with H
CUR.
Change the frequency as: 20Hz, 100Hz, 1kHz,
10kHz, 100kHz, 200kHz, 300kHz
(
STB8827
A is 300kHz
)
.
The reading of frequency meter should meet the demand of the test signal
frequency in this chapter.
6.6.6 Measurement accuracy
Basic parameters are R, L, C and D
,
so measurement accuracy is mainly about R,
L, C and D.
6.6.7 Accuracy of C and D
Function C
p
-D
Test frequency 100Hz 1kHz 10kHz 100kHz Test respectively
Level 1V
Range AUTO