STB8827
Series Operation Manual Chapter 7 Command Reference
93
NOTE: This command can be followed by a suffix unit A (ampere).
Query syntax: LIST:CURRent?
Return format: <NR3>[,<NR3>*]<NL^END>
NOTE: The current of each sweep point should be set from 50
μ
A to 20mA,
or return format will report errors.
The LIST:BIAS:VOLTage command is used to clear the original DC bias voltage of
each sweep point and reset the voltage. The LIST:BIAS:VOLTage? query returns
the current DC bias voltage of each sweep point.
Command syntax: LIST:BIAS:VOLTage<value>[,<value>*]
NOTE: * part means 201 sweep points at most can be set.
Where,
<value> is NR1, NR2 or NR3 data format.
For example: WrtCmd (“LIST:BIAS:VOLT 1.5V”)
Set the DC bias voltage of sweep point 1 as 1.5V.
Query syntax: LIST:BIAS
:
VOLTage?
Return format: <NR3> [, <NR3>*] <NL^END>
The LIST:BIAS:CURRent command is used to clear the original DC bias current of
each sweep point and reset them. The LIST:BIAS:CURRent? query returns the the
DC bias current of each sweep point.
Command syntax: LIST:BIAS:CURRent<value>[,<value>*]
NOTE: * part means 201 sweep points at most can be set.
Where,
<value> is NR1, NR2 or NR3 data format.
For example: WrtCmd (“LIST:BIAS:CURR 100MA”)
Set the DC bias current of the sweep point 1 as 100mA.
WrtCmd (“LIST:BIAS:CURR 1E-2, 2E-2, 3E-2,4E-2”)
Set the DC bias currents of sweep points 1, 2, 3 and 4 respectively
as 10mA, 20mA, 30mA and 40mA.
Query syntax: LIST:BIAS:CURRent?
Return format: <NR3>[,<NR3>*]<NL^END>
NOTE:
STB8827
series instrument has installed a 10V/100mA internal
DC bias current source. If 1A DC source is required, please install
an external source.
STB8827
can be used with TH1778 DC Bias
Source (providing DC current from 0 to 20A and 120A can be reached at
most).