7 / 9
2.3 Electrical connection
- Before connecting the unit, note the specifications on the name plate.
- The unit may be connected only by a certified electrician in compliance with country-specific and
local regulations.
- The unit is delivered with a connection cable of approx. 1.5 m without plug. If the terminal line has
to be replaced for structural reasons, a type H07BQ-F cable is to be used.
- An all-pole electric separator with a contact distance of at least 3 mm is to be connected in series
near the unit by the customer. The connection cable shall only be replaced or extended by the
manufacturer or an authorised Salvis Service.
- Each unit requires its own power connection and fuse.
- The unit must be included in the potential equalization system.
The M6 thread for the connection terminal is located behind the guide column
.
.
2.4 Dimensioned drawing of the table model
SA L301
130
600
490
550
600
40
0
772
372
80
from above
4 M6 threads
SA L300
450
227
273
40
10
40
360
480
772
292
80
60
90
15
17
5
37
2
205
60
70
95
30
65
450
500
80
512
204
162
name plate
Installation note:
Never carry the unit by the reflector!
Lay the cable upward neatly in a cable conduit.