CWUL-05/10/20 Series 

   

<3> Installation 

The chiller is designed for Ultraviolet Laser, it is easy to install and use. Please 

follow below steps for the first time use of new chiller.   

1.    Open  the  package  to  check  if  the  machine  is  intact  and  all  the  necessary 

accessories are completed. 

2.    Open the injection port and pour cooling water.Observing the water level gauge 

while pouring water, please do it slowly do not overflow. For cold area at north 

China, should add noncorrosive antifreeze fluid.   

3.    Well connect the water inlet and outlet pipes according to apply equipments.   

4.    Plug in power, turn on the power switch. (Do not turn on the machine without 

water in the water tank!) 

1

The  circulation  pump  is  starting work  after  the  power switch  turned on. The 

first  time  of  operating  may  cause  more  bubbles  in  the  pipe  leading  to  flow 

alarming occasionally, but keep it running for a few minutes later, it will back 

to normal later. 

2

For the first boot, will required immediate attention to check whether the water 

pipe leaks.   

3

It  is  normal  the  fan  and  other  components  of  machine  not  works  if  water 

temperature lower than set water after started the machine. The temperature 

controller  will  automatically  control  the  working  status  of  compressor, 

magnetic valve and fan etc components upon the set parameters.   

4

As it takes time for the compressor etc components to start, the waiting period 

could range from tens second to few minutes according to different working 

condition  before  the  machine  start,  so  do  not  turn  on  and  off  the  chiller 

frequently.   

5.    Check the water level of water tank   

The  water  level  in  water  tank  will  drop  slightly  as  the  air  in  water  pipe  was 

empty after new machine turned on. It is ok to refill certain amount of water in 

order  to  keep  the  water  level  around  green  area.  Observe  and  record  current 

water level, check the water level again after chiller running for a period of time, 

if water level drops obviously, will require to check whether the water pipeline 

leaks.     

6.    Parameters of temperature controller adjustment   

        Default setting of CWUL-05/10/20 series UV laser chiller temperature controller 

is 25

 water temperature cooling UV parameter, usually do not need to adjust 

the  control  parameter  ,but  if  special  require  for  specific  temperature  ,please 

refer P6 of 

Operation and Parameters adjustment

 

Summary of Contents for CWUL-05

Page 1: ...No 1 1 CWUL 05 10 20 Ultraviolet Laser Exclusive Chiller User Manual...

Page 2: ...2 Catalog 1 Cautions 3 2 Contour and parts induction 4 3 Installation 5 4 Operation and parameters adjustment 6 5 Alarm and output ports 10 6 Technical parameters 11 7 Water pump performance curve 14...

Page 3: ...odel according to actual circumstance 4 In order to protect the water pump it is strictly forbidden to run the chiller without water in water tank The new machine will empty all water in water tank be...

Page 4: ...ook and Components WATER INJECTION PORT COOLING WATER OUTLET COOLING WATER INLET AIR OUTLET DRAIN WATER LEVEL GAUGE ALARM OUTPUT TERMINAL POWER SOCKET WITH FUSE WITH FUSE POWER SWITCH TEMPERATURE CONT...

Page 5: ...of machine not works if water temperature lower than set water after started the machine The temperature controller will automatically control the working status of compressor magnetic valve and fan e...

Page 6: ...lve working Heating ON heating rod working INT Ctrl ON controller working in intelligent control mode CT Ctrl ON controller working in constant temperature control mode PARAM Set ON controller working...

Page 7: ...r will alarm the alarm code display E2 6 F5 Ultralow water temperature alarm 1 20 15 0 When water temperature low to F0 F4 the chiller will alarm the alarm code display E3 7 F6 Ultrahigh ambient tempe...

Page 8: ...ccording to ambient temperature changes The temperature difference is controlled by F1 6 Advanced parameters adjustment cases 1 Case 1 26 8 cooling UV laser settings Chiller under constant temperature...

Page 9: ...rature display without saving the modified parameters Order Code Content Case 1 Set value Case 2 Set value Case 3 Set value T 506B Temperature controller default setting 1 F0 Temperature setting 26 8...

Page 10: ...ss than 300V Display Condition Alarm code Buzzer OUT H1 H2 OUT H1 H3 Circulating pump works properly DISCONNECTION BREAKOVER Ultrahigh room temp E1 Sounds BREAKOVER DISCONNECTION Ultrahigh water temp...

Page 11: ...er temperature alarm Pump power 50 W Inlet and outlet External 10mm barbed connector Max lift 12M Max flow 13L min Tank capacity 6 L N W 23Kgs G W 26Kgs Dimension 58X29X47 cm L X W X H Package dimensi...

Page 12: ...pillary Protection Overcurrent protection for compressor flow alarm over temperature alarm Pump power 50 W 100 W Max lift 12M 25M Max flow 13L min 16L min Tank capacity 6 L N W 24Kgs G W 27Kgs Dimensi...

Page 13: ...Capillary Protection Overcurrent protection for compressor flow alarm over temperature alarm Pump power 50 W 100 W Max lift 12M 25M Max flow 13L min 16L min Tank capacity 6 L N W 26Kgs G W 29Kgs Dime...

Page 14: ...CWUL 05 10 20 Series 14 7 Refrigeration performance curve Performance curve of chiller under ambient temperature 20 30 and 40...

Page 15: ...are blocked or a pipe bending deformation Check water circulation pipe Ultrahigh water temperature alarm controller displays E2 Blocked dust gauze bad thermolysis Unpick and wash the dust gauze regula...

Reviews: