Troubleshooting
Insufficient airflow
If airflow through the ozonizer drops below
75 litres per hour, ozone production will shut
down to prevent overheating. Please provide
sufficient airflow to the ozonizer, removing any
blockages.
High temperature shutdown
If there is insufficient ventilation for cooling or
insufficient airflow, both combined with high-
level ozone production, the temperature in the
ozonizer may exceed its limit (70 °C). In this
event, the ozonizer will shut down to prevent
damage. Please provide sufficient circulation of
air, in particular if the ozonizer is mounted in
a cabinet. The ozonizer will resume operation
once the temperature has dropped to be within
limits.
High voltage malfunction
Water or dirt particles entering the ozone
generator may cause damage, creating a faulty
voltage potential inside the ozonizer. In this
event, please disconnect the ozonizer from
the power supply immediately, removing the
wall plug from the mains socket. Blowing air
through the ozonizer may remove any water
that has entered. Please cleanse the ozone
generator as described above. If the problem
persists, please send the ozonizer in for
inspection at Sander company headquarters
(see address on back cover page).
Disposal
Packaging is made from eco-friendly
materials that may be disposed of at
any local recycling facilities.
Should you need to dispose of the
ozonizer, please do not place it into
your regular domestic waste. In the
interest of environmental protection,
please use an appropriate disposal facility for
electrical equipment. Your local (municipal)
government will provide you with the locations
and opening times of such facilities.
Specifications
⇒
Ozone output:
2000 milligrams per hour
⇒
Power supply:
100–240 volts AC (50/60 Hz)
⇒
Current consumption:
1,5 ampères
⇒
Operating voltage:
24 volts DC