7
INSTALLATION
F
IXTURE
P
LACEMENT
The
Sanitaire
®
Model RSCS280 fixture is recommended for use in rooms with suspended "T-Bar" ceilings.
Access to the room does not need to be controlled while the fixture is operating. The exposure chamber, of the
Sanitaire
®
fixture, protects the room's occupants from ultraviolet exposure by restricting ultraviolet radiation from
passing into the room.
To ensure safe and proper use;
•
The
Sanitaire
®
Model RSCS280 Room Air Sanitizer should be installed by qualified professionals so that local
and national building and electrical codes are fully complied with.
WARNING
: Failure to adequately support this equipment could result in serious injury or even death.
•
The
Sanitaire
®
Model RSCS280 has 4 suspension points. The suspension method chosen should be structurally
adequate to support the weight of the unit independently from any one of these suspension points (unit weight can
be found in the equipment specification).
•
Avoid using a single cable or chain or other structural element to support more than one suspension point.
•
Never attempt to support or suspend this unit on the ceiling tile track system.
•
Remember it is the equipment user’s responsibility to insure an adequate and safe installation of this equipment
and the safety of room occupants.
•
It is recommended that the
Sanitaire
®
Model RSCS28 be wired to a wall switch to control on and off operation.
!
OPERATION AND MONITORING CONTROLS
The
Sanitaire
®
is designed to operate with minimal operator supervision. Fixture operation is monitored from the Base Panel.
Lamp Operation Indicators:
Provide positive indication, when lit, of each germicidal lamp operation. Lamp
Operation LEDs, which are not lit, indicate faulty germicidal lamps or ballasts. Lamp Operation LED position, in the
LED array, is relative to the germicidal lamp position in the exposure chamber, aiding in the location of the faulty
lamps or ballast(s).