Sanosil Q-Jet Advanced
18
5.5
Step (max. 3 cycles)
The complete disinfection process can be split in different steps
The disinfection result (efficacy) will be higher when done in cycles (up to 100%)
Please read and follow important safety directions on page 6
before use!
•
Choose STEP and confirm by pressing ENTER
•
Up to 3 steps can be set (CYCLE 1-3) as follows:
ON =
cycle’s
run-time in min. 0-59 = active fogging time with a dosage of 33ml/min
OFF =
cycle’s
interval time in min. 0-59 = no fogging, activation and dwell time
•
Press ENTER to confirm each setting!
•
To get to the different cycles, press every time ENTER
•
Each cycle has an ON- and an OFF- time. Please input for
each cycle the desired time (between 0 and 59 minutes) and
confirm with ENTER. Enter
“
0
”
on cycle 3 if not used.
WARNING
The dosing values below/on the next page (19) are a recommendation for well pre-cleaned
rooms only and are based on many years of experience and are partly supported by test
reports. However, the values given are not binding as the disinfection effect can be influenced
by many factors. Proof of efficacy must be confirmed by microbiological checks after
disinfection.
For room treatments according norm Afnor T72-281 (2014) refer to the table on page 19
CYCLE 1
ON: 1m
OFF: 1m
CYCLE 2
ON: 1m
OFF: 1m
STEP
CYCLE 3
ON: 1m
OFF: 1m