Sanosil Q-Jet Compact
13
4.5.3
Procedure before re-entering the room
•
Check with a hydrogen peroxide level sensor that the H
2
O
2
-Level is below 1ppm
•
Open windows and doors for min. 15 minutes, unseal HVAC vents and restart HVAC system to allow increased airflow
circulation in the room. Once a safe ppm level is reached, room can be re-used again.
WARNING!
Risk of injuries!
To prevent injuries do always check H
2
O
2
-Level before entering the room.
When entering room to measure the H
2
O
2
-Value, do always wear full protection equipment,
i.e. protection mask, safety googles and gloves.
The disinfected room can only be entered without protection equipment if the H
2
O
2
-Value is below the local
maximum
permissible
value specified by the authorities.
•
Tip ON/OFF (3) to switch off device
INDICATION
For transportation as well as for longer storage the liquid inside the device has to be completely emptied.
(please see chapter 4.5.4)
4.5.4
Emptying the Tank
CAUTION!
Risk of injuries!
Do always wear gloves and safety googles when emptying the tank!
•
Before emptying the tank always turn OFF the device and
disconnect the power supply plug!
•
Unscrew rear panel and remove it
•
Pull drain hose to the outside, remove tap and open valve to
allow the release of residual liquid!
•
Do completely empty the device by tilting it slightly towards
the drain connection of the tank!
•
Do close the valve after emptying, put back the tap and stow the
hose inside the device.
•
Fix the rear panel.