Sanosil Q-Jet Compact 

 

 

 

 
 
 

2.9. Important safety Directions 

(Read carefully before use) 

 
 

Caution (atomized Sanosil Disinfectant): 

 

1.

 

When the Q-Jet has completed its dispensing cycle, the room will be full of atomized  Sanosil Disinfectant. The 
treated  room  MUST  REMAIN  UNOCCUPIED  FOR  AT  LEAST  120  MINUTES  AFTER  FOGGING.  120  minutes  is  the 
minimum wait time and means not that the room is safe to re-enter or reoccupy. Length of wait time increases 
with fogging quantity (ml/m

3

). 

 

2.

 

When room re-entry is required before treatment process is completed or to perform a H

2

O

2

 level check , 

proper personal equipment (PPE) including wraparound style goggles; respirator mask (filter class ABE1) 
filtering hydrogen peroxide (for example: 3M 4277 or 3M 7500 mask with 6057 filter) and gloves should be 
worn to safely enter the room. 

 
 

Caution (Sanosil Q-JET device) 

 
A caution statement in this manual identifies a condition or practice, which if not corrected or discontinued immediately, 
could lead to serious injury or death. 
 

1. 

KEEP OUT OF REACH FOR CHILDREN. 

 

2. 

Carefully read all instructions, warning, caution and first aid statements prior to use. 

 

4. 

Use only as directed. 

 

5. 

The device is designed for easy-use. However, only properly trained individuals should use this device. 

 

6. 

Do not leave the device unattended between uses. 

 

7. 

Electric shock hazard. Do not plug in if electrical plug or parts are wet. 

 

8. 

Use only a Sanosil supplied electrical power cord. 

 

9. 

Take care to prevent water or other fluid from entering the device. Should this occur, allow to completely dry before 
use. Check the accuracy of all operating functions. 

 

10.

 

Always unplug the power cord from the outlet before moving the device. 

 

11.

 

There are no user-serviceable parts inside the device other than the nozzle. An operator may only perform 
maintenance procedures specifically described in this manual, all other repairs should be made by the manufacturer. 

 

12.

 

Position the device on a secure surface to prevent rolling or moving while in use. 

 

13.

 

Do not tip the device on its side at any time unless completely empty and dry. 
as this can cause leakage and damage on electrical parts 

 

14.

 

Do not overfill the tank as this can cause leakage and damage on electrical parts 

 

15.

 

Protect this device from severe impact and shock. 

 

16.

 

While in use, the body of the device may become warm to the touch. Handle with caution.  

 

17.

 

Read the specific warning and caution statements of the Sanosil disinfectant you use. 

 

18.

 

To ensure effectiveness and safety, use 

only

 

with Sanosil disinfectant. Using any other manufacturer’s disinfecting or 

cleaning product will result in serious injury, exposure and environmental damage. 

 

19.

 

Sanosil Ltd. is not responsible for any injury or damage caused by using this device outside of the specific parameters 
detailed in this manual. 

 

20.

 

Store in a safe, dry location. Do no place anything on top of the device. Store in an upright position. 

 

 

 

Summary of Contents for Q-Jet Compact

Page 1: ...Sanosil Q Jet Compact 1 Operation and Maintenance Manual Sanosil Q Jet Compact Des M 56en Compact V2 English...

Page 2: ...8 3 1 Technical Data 8 3 2 Parts and Functions 8 3 3 Package Contents 8 3 4 Q Jet Compact Accessories 8 3 5 Spare Parts 8 3 6 Recommended Equipment 8 3 7 Trolley optional 9 4 Getting Started and Oper...

Page 3: ...ries Low risk Disregard can lead to medium or light injuries Disregard can damage thedevice Gives good advice for operating thedevice Further symbols will be explained thereafter 2 Safety DANGER Instr...

Page 4: ...Sanosil Q Jet Compact 4 2 1 CE Declaration of Conformity manufacturer...

Page 5: ...le liquids Do always and only use Sanosil disinfection liquids 2 4 Danger and Caution Warnings The device must be used in full compliance with this Operation and Maintenance Manual It is strictly forb...

Page 6: ...an as needed basis depending on the volume of traffic and use Users should familiarize themselves with Q Jet operation by reading the Q Jet User Manual prior to first use It is essential that Q Jet u...

Page 7: ...rical plug or parts are wet 8 Use only a Sanosil supplied electrical power cord 9 Take care to prevent water or other fluid from entering the device Should this occur allow to completely dry before us...

Page 8: ...ent 0 20ppm for example Dr ger X am 5100 Gloves 3 Product Description 3 1 Technical Data Powerrequirements 220 240 V 50Hz Amperage max 5 A Operatingtemperature 5 C to 40 C Engine Power 800 W Engine ro...

Page 9: ...place the device on top of trolley secure the device with the screw on the trolley WARNING Risk of crushing Hands and fingers can be crushed between device and trolley Do always wear device on handle...

Page 10: ...ctrical connections carefully check that the voltage supplied is compatible with the voltage marked on the dataplate Insert power cord 2 in IEC power socket 4 3 Filling Remove tank cap 6 Fill disinfec...

Page 11: ...x Width x Height of the room in meters and round it up to the next cubic meter Check relative humidity should not be lower than 50 Establish nebulization time room s volume x desired concentration 30...

Page 12: ...the waiting time the room can be prepared for re entry please see 4 5 3 Shortening of waiting time after a minimal waiting time of 120 minutes place a strong dehumidifier in the middle of the room and...

Page 13: ...n equipment if the H2 O2 Value isbelow the local maximum permissible value specified by the authorities Tip ON OFF 3 to switch off device INDICATION For transportation as well as for longer storage th...

Page 14: ...eneral Visual control Checking operationality daily Components status of device s components integrity of components fluid loss sealings caps monthly Documentation Verification that all documentation...

Page 15: ...gine temperature Air intake filter toodirty Check engine switch off and on Set up device once more Clean or change filter Liquid is not sprayed Tank is empty Nozzle is clogged Fill liquid Remove obstr...